Standard Operating Procedure (SOP) for HPLC Calibration
Objective
To outline the procedure for the calibration of High-Performance Liquid Chromatography (HPLC) equipment to ensure accurate flow rate, injector precision, detector linearity, and column oven temperature, meeting specified criteria.
Scope
This procedure applies to the calibration of HPLC systems used in analytical laboratories, specifically for:
- Accuracy of Flow Rate: 0.5 mL/min and 5 mL/min.
- Injector Precision: 20 μL injection of 50.0 μg/mL caffeine solution.
- Detector Linearity: Wavelength at 273 nm.
- Column Oven Temperature: Set at 40°C.
Responsibilities
Analyst:
- Perform calibration as per the procedure.
- Record observations in the calibration logbook.
Quality Assurance (QA):
- Verify calibration records and ensure adherence to the SOP.
- Approve calibration certificates.
Materials and Equipment
Materials:
- Caffeine reference standard (purity ≥ 99%).
- HPLC-grade water, acetonitrile, and methanol.
- Mobile phase specific to the analysis method.
Equipment:
- HPLC system (with flow rate, injector, detector, and oven components).
- Calibrated analytical balance.
- Volumetric flasks and pipettes (calibrated).
- Stopwatch or timer.
Procedure
1. Accuracy of Flow Rate
Purpose:
To verify the flow rate accuracy at 0.5 mL/min and 5 mL/min.
Steps:
- Set the HPLC pump flow rate to 0.5 mL/min using the control panel or software.
- Collect the mobile phase exiting the column outlet into a pre-weighed container for 5 minutes.
- Weigh the collected liquid and calculate the flow rate using the formula:Flow Rate (mL/min)=Weight of collected liquid (g)Time (min)×Density (g/mL)\text{Flow Rate (mL/min)} = \frac{\text{Weight of collected liquid (g)}}{\text{Time (min)} \times \text{Density (g/mL)}}
(Assume the density of water = 1 g/mL if used as the mobile phase.)
- Repeat steps 1-3 for a flow rate of 5 mL/min.
- Record results and ensure deviation from the set flow rate does not exceed ±1%.
2. Injector Precision
Purpose:
To evaluate the precision of injections at 20 μL.
Steps:
- Prepare a caffeine solution of 50.0 μg/mL in HPLC-grade water.
- Set up the HPLC system with the specified mobile phase and column.
- Inject the prepared caffeine solution six times (20 μL per injection).
- Record the peak area for each injection.
- Calculate the %RSD (Relative Standard Deviation) using the formula:%RSD=(Standard Deviation of Peak AreasMean Peak Area)×100\%\text{RSD} = \left( \frac{\text{Standard Deviation of Peak Areas}}{\text{Mean Peak Area}} \right) \times 100
- Ensure %RSD does not exceed 2%.
3. Detector Linearity
Purpose:
To assess detector response linearity at a wavelength of 273 nm.
Steps:
- Prepare caffeine standard solutions with concentrations of 10, 20, 30, 40, and 50 μg/mL.
- Set the detector wavelength to 273 nm.
- Inject each standard solution in triplicate and record the peak areas.
- Plot a calibration curve of concentration (x-axis) vs. peak area (y-axis).
- Calculate the correlation coefficient (R²). Ensure R² ≥ 0.999.
4. Column Oven Temperature
Purpose:
To verify the accuracy of the column oven temperature set at 40°C.
Steps:
- Set the column oven temperature to 40°C.
- Place a calibrated external thermometer or temperature probe inside the oven.
- Allow the oven to stabilize for 30 minutes.
- Record the temperature reading from the thermometer.
- Ensure the deviation between the set temperature and measured temperature does not exceed ±0.5°C.
Acceptance Criteria
Calibration Parameter | Acceptance Criteria |
---|---|
Flow Rate Accuracy | ±1% of set flow rate |
Injector Precision | %RSD ≤ 2% |
Detector Linearity | R² ≥ 0.999 |
Column Oven Temperature | Deviation ≤ ±0.5°C |
Documentation
- Record all calibration data in the HPLC Calibration Logbook.
- Attach calibration certificates and raw data (e.g., chromatograms).
- Submit the calibration report to QA for approval.
References
- Manufacturer’s manual for the HPLC system.
- ISO 9001:2015 – Quality Management Systems.
- USP <621> Chromatography.
Zulkifal
0
Tags :