Standard Operating Procedure (SOP) For Autoclave Hirayama HVE-50
Objective
To provide a comprehensive guide for the operation, maintenance, and troubleshooting of the Autoclave Hirayama HVE-50, ensuring effective sterilization of media, glassware, laboratory equipment, and waste decontamination.
Scope
This SOP is applicable to all sterilization activities using the Autoclave Hirayama HVE-50, including media preparation, glassware sterilization, lab equipment sterilization, and waste decontamination.
Responsibilities
Roles and Tasks
- Analyst/User/Technical Staff
- Perform tasks as outlined in the SOP.
- Maintain and record all activities per the SOP.
- Analyst/DTM/TM/DQMSM
- Draft and update the SOP to ensure compliance with Good Laboratory Practices (GLP).
- QMS Manager
- Review and ensure SOP adherence to prescribed formats.
- Monitor implementation of SOP-related activities.
- Technical Manager
- Supervise technical procedures in line with the SOP.
- Head Of Organization
- Authorize the SOP and verify activity implementation.
Terms & Definitions
- SOP: Standard Operating Procedure
- QMS: Quality Management System
- DQMSM: Deputy QMS Manager
- GLP: Good Laboratory Practices
- PSI: Pounds per Square Inch
- Calibration: Aligning measurement equipment to standard benchmarks.
- Validation: Assessing the performance of measurement equipment.
Procedure
Responsibilities and Tasks
- Installation: Service Engineer/Vendor
- Transportation: DMO/BME
- External Calibration: Vendor
- Preventive Maintenance: Analyst/DMO/BME
- Service: Service Engineer/Vendor
- Autoclaving: Analyst/Senior Microbiologist/Assistant Bacteriologist
Operating Procedure
- Powering On:
- Turn ON the breaker.
- Press the power ON/OFF switch.
- Water Pouring:
- Ensure the water level is sufficient for the sterilization cycle.
- Loading Items:
- Place items in the chamber, leaving room for steam circulation.
- Setting Parameters:
- Adjust temperature (121°C), pressure (15 PSI), and time (5 minutes).
- Use Mode 1 for default settings, including warming at 50°C and natural cooling (P-0).
- Starting the Cycle:
- Lock the lid and press the start button.
- Monitor the sterilization cycle on the display.
- Completion:
- Allow temperature to drop to 50-65°C before unloading.
- Loosen the pressure nozzle, if necessary.
- Unloading:
- Unlock and open the lid.
- Use insulated gloves to remove items.
Precautions
- Maintain the exhaust water level between LOW and HIGH.
- Avoid overloading to ensure proper steam circulation.
- Do not open the chamber until the pressure and temperature are safe.
- Use heat-resistant gloves for handling hot items.
- Allow items to cool before transportation.
Troubleshooting
Error Code | Description |
---|---|
Error:1 | Lack of water alarm |
Error:2 | Temperature sensor wire breakage |
Error:3 | Over-temperature alarm |
Error:4 | Over-cool alarm |
Error:5 | Over-pressure alarm |
Error:6 | Lid open alarm |
Error:7 | Automatic exhaust valve trouble alarm |
Error:9 | Sterilization heater trouble alarm |
Error:L | Open/close lever locking failure |
Error:E | Exhaust bottle anomaly alarm |
For detailed troubleshooting, consult the instrument manual.
Verification
- Use steri tape or steri gauge to verify sterilization effectiveness.
- Ensure visible color change to confirm sterilization.
Calibration
- Perform annual external calibration to maintain accuracy.