Standard Operating Procedure (SOP) For Autoclave Hirayama HVE-50

Objective

To provide a comprehensive guide for the operation, maintenance, and troubleshooting of the Autoclave Hirayama HVE-50, ensuring effective sterilization of media, glassware, laboratory equipment, and waste decontamination.

Scope

This SOP is applicable to all sterilization activities using the Autoclave Hirayama HVE-50, including media preparation, glassware sterilization, lab equipment sterilization, and waste decontamination.

Responsibilities

Roles and Tasks

  • Analyst/User/Technical Staff
    • Perform tasks as outlined in the SOP.
    • Maintain and record all activities per the SOP.
  • Analyst/DTM/TM/DQMSM
    • Draft and update the SOP to ensure compliance with Good Laboratory Practices (GLP).
  • QMS Manager
    • Review and ensure SOP adherence to prescribed formats.
    • Monitor implementation of SOP-related activities.
  • Technical Manager
    • Supervise technical procedures in line with the SOP.
  • Head Of Organization
    • Authorize the SOP and verify activity implementation.

Terms & Definitions

  • SOP: Standard Operating Procedure
  • QMS: Quality Management System
  • DQMSM: Deputy QMS Manager
  • GLP: Good Laboratory Practices
  • PSI: Pounds per Square Inch
  • Calibration: Aligning measurement equipment to standard benchmarks.
  • Validation: Assessing the performance of measurement equipment.

Procedure

Responsibilities and Tasks

  • Installation: Service Engineer/Vendor
  • Transportation: DMO/BME
  • External Calibration: Vendor
  • Preventive Maintenance: Analyst/DMO/BME
  • Service: Service Engineer/Vendor
  • Autoclaving: Analyst/Senior Microbiologist/Assistant Bacteriologist

Operating Procedure

  1. Powering On:
    • Turn ON the breaker.
    • Press the power ON/OFF switch.
  2. Water Pouring:
    • Ensure the water level is sufficient for the sterilization cycle.
  3. Loading Items:
    • Place items in the chamber, leaving room for steam circulation.
  4. Setting Parameters:
    • Adjust temperature (121°C), pressure (15 PSI), and time (5 minutes).
    • Use Mode 1 for default settings, including warming at 50°C and natural cooling (P-0).
  5. Starting the Cycle:
    • Lock the lid and press the start button.
    • Monitor the sterilization cycle on the display.
  6. Completion:
    • Allow temperature to drop to 50-65°C before unloading.
    • Loosen the pressure nozzle, if necessary.
  7. Unloading:
    • Unlock and open the lid.
    • Use insulated gloves to remove items.

Precautions

  • Maintain the exhaust water level between LOW and HIGH.
  • Avoid overloading to ensure proper steam circulation.
  • Do not open the chamber until the pressure and temperature are safe.
  • Use heat-resistant gloves for handling hot items.
  • Allow items to cool before transportation.

Troubleshooting

Error Code Description
Error:1 Lack of water alarm
Error:2 Temperature sensor wire breakage
Error:3 Over-temperature alarm
Error:4 Over-cool alarm
Error:5 Over-pressure alarm
Error:6 Lid open alarm
Error:7 Automatic exhaust valve trouble alarm
Error:9 Sterilization heater trouble alarm
Error:L Open/close lever locking failure
Error:E Exhaust bottle anomaly alarm

For detailed troubleshooting, consult the instrument manual.

Verification

  • Use steri tape or steri gauge to verify sterilization effectiveness.
  • Ensure visible color change to confirm sterilization.

Calibration

  • Perform annual external calibration to maintain accuracy.