SOP For Analytical Weighing Balances RADWAG / AS 220/C/2
Scope
This SOP applies to the RADWAG / AS 220/C/2 analytical electric weighing balances for accurate and reliable measurements.
Responsibilities
Roles and Tasks
- Analyst/User/Technical Staff
- Perform all activities as per the SOP.
- Maintain and record activities according to the SOP.
- Analyst/DTM/TM/DQMSM
- Draft and update the SOP to enhance compliance with Good Laboratory Practices (GLP).
- QMS Manager
- Review and verify the SOP format.
- Ensure SOP implementation for all activities.
- Technical Manager
- Supervise technical procedures as per the SOP.
- Head Of Organization
- Authorize the SOP and verify its implementation.
Terms and Definitions
- SOP: Standard Operating Procedure
- QMS: Quality Management System
- GLP: Good Laboratory Practices
- Calibration: Aligning equipment to a traceable standard under controlled conditions.
- Validation: Checking equipment performance to ensure accuracy.
Procedure
Installation
- Power Requirements:
- Select a site near a power source. Ensure the voltage conforms to the AC adapter specifications.
- Installation Site:
- Ensure the site is level, stable, and free from air currents, vibrations, sunlight, temperature fluctuations, or electromagnetic interference.
- Adjust the balance bubble to the center using the leveling screws.
Calibration and Verification
- Calibration:
- Perform annual calibration or when:
- Installation site changes.
- Abnormal behavior is detected during weighing.
- Use traceable standard weights calibrated by an ISO 17025:2005 accredited lab.
- Perform annual calibration or when:
- Daily Verification:
- Conduct by trained personnel.
- Record the temperature and humidity before and after verification.
- Follow these steps:
- Turn on the balance and allow it to self-check.
- Verify using standard weights (10 mg, 100 mg, 1000 mg, and 10 g).
- Record results in the logbook or attach printouts.
Operating Procedure
- Ensure daily verification is complete.
- Verify the bubble is centered and the area is clean.
- Press [O/T] to reset the balance to 0.0000 g.
- Place a butter paper or weighing boat on the pan and press [O/T].
- Add the item to be weighed and note the stable reading or press the ‘Print’ key.
Preventive Maintenance and Cleaning
- Before Use:
- Ensure the bubble is centered.
- Avoid working near air currents or vibrations.
- During Use:
- Close the glass doors when not placing or removing items.
- Use dry spatulas and containers for weighing.
- After Use:
- Clean spills immediately using a damp cloth.
- Remove the weighing pan for thorough cleaning.
- Transport:
- Avoid moving the balance without proper authorization.
Precautions
- Perform all activities as per the SOP.
- Ensure only trained personnel handle the balance.
- Place the balance on a stable, flat surface.
- Check the bubble before use.
- Record all activities diligently.
Calibration Schedule
- External Calibration: Annually
- Daily Verification: Before use
Related Documents
(Include references if available)
References
(Include citations if applicable)